AID10 Remote Alarm Indicator

● Remote monitoring
● LED indicators: On, Insulation, Overload, Overtemp
● Mute button
● Test button
● Auxiliary power supply: DC24V
● RS485(MODBUS-RTU)

description1
Zipper closure 1/5 zip athletic pullovers for men. Stretchy, lightweight, fast-drying fabric for superior performance. REGULAR FIT - US standard sizes. An athletic fit that sits close to the body for a wide range of motion, designed for optimal performance and all day comfort. FEATURES - Quarter zip closure;Thumbholes on long sleeves to keep them in place during workout
product detail

General

AID series alarm and display device is a remote display and sound and light alarm device by using advanced fieldbus communications technology. The device can be installed on the operating room or ICU, the medical staff can easily know the status of medical isolated power systems, and the type of fault when the system fails.

Function

- The insulation resistance alarm value, load current alarm value and transformer temperature alarm value of the system can be set remotely.

- When the system appears insulation fault, overload, transformer temperature over limit and wiring fault, alarm and display instrument gives out the corresponding sound and light alarm, and has the function of eliminating sound alarm.

- Using advanced fieldbus technology, the remote monitoring function can be realized by real-time data interaction with the insulation monitor.

- It can monitor one set of AIM-M100 insulation monitor and be used for installation by embedding into wall.

- It can be applied to monitor operating room or ICU only using single set of isolated power.

Wiring & Installation

Dimension

    • Front view
    • Side view

    Product Selection


     

    Structure


     

    LED Indicator Instructions

    Indicator light

    Instructions

    On

    When the instrument operation is normal, the indicator flashes, with the flashing frequency of about one time per second.

    Insulation

    When the insulation resistance exceeds the alarm value, the indicator flashes to alarm.

    Overload

    When the load current exceeds the rated current of the transformer, the indicator flashes to alarm.

    Overtemp

    When transformer temperature exceeds alarm value, indicator flashes to alarm.

    Button Function Description

    Button Function Description
    Mute When there is alarm, press this button to eliminate the alarm sound.
    Test Used to start the self-test function of monitor.

     

    Common Problem

    For common problems, the causes can be determined and the faults can be eliminated according to the phenomenon of each instrument and the fault types:

    Fault phenomenon

    Possible Causes and Troubleshooting

    Abnormal communication

    1) The address of AIM-M10 is not set to default 1, or BUAD is not set to default 9600, so it should be set to default value.

    2) The communication line with AIM-M10 is not well connected. Check the communication line and confirm that the matching resistance is well connected.

    PARAMETER

    Parameter               Type

    AID10

    AID150

    Auxiliary power supply

    Voltage

    DC 24V

    Consumption

    < 0.6W

    Insulation alarming range

    ——

    14A, 18A, 22A, 28A, 35A, 45A

    Temperature alarm setting range

    ——

    0~+200

    Number of monitored systems

    1

    16

    Alarm method

    Sound-light alarm

    Alarm type

    Insulation failure, overload, overheat, equipment failure

    Communication mode

    RS485, MODBUS-RTU

    Display mode

    LED display

    128×64  LCD display