AID150 Remote Alarm Indicator

● Remote monitoring
● LED indicators: On, Insulation, Overload, Overtemp
● Mute button
● Test button
● Auxiliary power supply: DC24V
● RS485(MODBUS-RTU)
description1
Zipper closure 1/5 zip athletic pullovers for men. Stretchy, lightweight, fast-drying fabric for superior performance. REGULAR FIT - US standard sizes. An athletic fit that sits close to the body for a wide range of motion, designed for optimal performance and all day comfort. FEATURES - Quarter zip closure;Thumbholes on long sleeves to keep them in place during workout
product detail

General

AID Series alarm and display device is remote display and sound and light alarm device by using advanced fieldbus communications technology. AID Series alarm and display device have four models: AID10, AID120, AID150 and AID200. Products are classified according to the installation method and display. The device can be installed on the operating room or information control panel of intensive care unit, the medical staff can easily know the status of medical isolated power systems, and the type of fault when the system fails.

AID150 centralized alarm and display instrument adopts the LCD liquid crystal display and achieves data exchange with AIM-M200 medical intelligent insulation monitoring instrument through CAN communication interface of self-defined protocol, which can real-time monitor multi-channel data of AIM-M200 medical intelligent insulation monitoring instrument.

 

Function

* Can read and display monitoring data and alarm information of insulation monitor in real-time by bus communication

* Can set the alarm parameters of insulation monitor remotely

* When the system fails, products can generate sound and light alarm, if you press the mute button, the sound can be turned off, but the corresponding fault indicator does not go out until the fault is removed so far.

 

Specification

Model

Parameter

AID10

AID120

AID150

AID200

Auxiliary power

Voltage

DC24V

Consumption

0.6W

Insulation resistance Measurement range

--

0-999KΩ

--

--

Insulation alarm value range

--

50-999KΩ

--

--

Transformer load rate display

--

Percentage display

--

--

Load current alarm value settings

--

14A,18A, 22A, 28A, 35A, 45A

Temperature alarm value range

--

0-200℃

Alarm

Sound and light alarm

Alarm type

Insulation fault, overload, over-temperature, equipment failure

Communication

RS485 interface, Modbus-RTU protocol

Display

--

LED display

LCD display

LCD display

 

Dimension

Display

 

Application

Wiring and Installation

Terminal Description

    Note: The 24V power supply can be connected by multiple copper wires of 2x1.5mm2, and the RS485 communication terminal can be connected by shielded twisted pair of 2x1.5mm2

    Installation

     

    Common Problems

    For common problems, the causes can be determined and the faults can be eliminated according to the phenomenon of each instrument and the fault types:

    Fault phenomenon

    Possible causes and troubleshooting

    The instrument is not lit.

    The 24V power supply is not connected well. Check the wirings of 24V and G terminals and rewire.

    Communication indicator does not flash

    In the system, the AIM-M monitor communication line is not connected properly. Check the communication line and confirm whether the matching resistance is connected properly.

     

    Note: If the above faults occur, interrupt the power to troubleshoot, and adjust the wirings until everything is normal

    SELECTION Table

    Acrel Medical IT System Insulation Monitoring and Fault Location Products Selection Table V1.0

    Note“□”is optional“—”is not optional“■”is selection.

    Medical Single Phase Isolation Transformer

    AITR-3150

    AITR-5000

    AITR-6300

    AITR-8000

    AITR-10000

    Capacity

    3150VA

    5000VA

    6300VA

    8000VA

    10000VA

    Insulation monitor

    AIM-M10

    AIM-M100

    AIM-M200

    Power supply

    AC220±10%

    AC85~265V

    DC100-300V

    DC24V

    Communication

    RS485

    RS485+RS485

    RS485+CAN

    Output

    DO1(over temperature)

    DO1(over temperature)

    DO2(alarm)

    DO1(over temperature)

    DO2(alarm) DO3(programmable)

    1 channel DC24V

    Support fault location

    No

    Yes

    Power supply

    ACLP10-24

    HDR-60-24

    Insulation monitor

    AIM-M100

    AIM-M200

    Alarm indicator

    AID10

    AID120

    AID150

    AID200

    Display

    LED

    LCD

    Multi-machine display

    No

    Yes

    Current transformer

    AKH-0.66P26

    Test signal generator

    ASG150

    Support fault location

    No

    Yes

    Fault locator

    AIL150-4

    AIL150-8

    Support fault location

    No

    Yes

    Locate number of channels

    <=4

    >4; <8

    >8 Note 1

    Other

    Note: 1. More than 8 channels can be freely combined, and the maximum support 24 locate channels.

     

    Ordering example

    Example 1: Isolation transformer 8kVA; AC220V; Insulation monitoring; Operating room LCD display; DO output programmable,According to the selection table, confirm the model is AITR-8000, AIM-M100, ACLP10-24, AKH-0.66P26, AID150.
    Example 2: Isolation transformer 10kVA; Insulation monitoring; Fault location; ICU LCD display; DO output; Locate 12 loops;
    According to the selection table, confirm the model is AITR-10000, AIM-M200, HDR-60-24, AKH-0.66P26, AID150, ASG150, AIL150-4, AIL150-8